T R I G G E R E D S W E E P O S C I L L O S C O P E C S - 1 3 5 2 DUAL T R A C E OSCILLOSCOPE I N S T R U C T I O N MANUAL...
Page 2
F E A T U R E S * Time base switch allows changeover between * Vertical axis provides high sensitivity and wide CHOP and A L T and between V (vertical) and H band width, 2 mV/div, 15 MHz ( - 3 dB). (horizontal) T V sync separator...
Page 3
S P E C I F I C A T I O N S Linearity: T y p e of Cathod R a y T u b e : C 3 3 1 P 3 1 B Better than 3 % (2)us/div ~ 0 5 s / d i v ) Acceleration Voltage: Better than 5% (0.5jus/div ~...
Page 4
Dimensions and Weight Input I m p e d a n c e : 10kfi ± 2 0 % Width: Frequency R e s p o n s e : 2 1 0 mm DC ~ 1 MHz Height: M a x i m u m Input A l l o w a b l e Voltage: 136 mm 50V (DC + A C peak)
C O N T R O L S ON P A N E L S FRONT PANEL Fig. S I D E PANEL Fig. 2...
Page 6
7. V O L T S / D I V 1 . $ P O S I T I O N Vertical attenuator Channel which Vertical position adjustment for CH1 trace and Y position for X - Y operation. Waveforms can provides step adjustment of vertical sensitivi- ty.
Page 7
1 5 . F O C U S slightly decreased. Spot focus control obtain optimum 2 1 . S W E E P T I M E / D I V waveform according to brightness. Horizontal coarse s w e e p time selector.
33. C O R D R E E L External DC power jack to supply 11 ~ 1 5.5V DC power. Used to wind power cord when the os- cilloscope is to be carried or stored. It also 32. V O L T A G E N A M E P L A T E serves as a stand when the oscilloscope is Be sure to use voltage and fuse specified.
difference between C H 1 and CH2 is displayed sure that the red lead is connected to " + " and (CH1 - CH2). the black lead to " — " side. 5. When the signal voltage is more than 2 m V and * After the batteries have been connected, insert waveform fails to appear on the screen, the os- the battery case from the rear and replace the...
Page 10
nector plug must be disconnected while the battery B E Z E L is being charged. H O O D When the battery is almost fully charged, the C H A R G E lamp becomes dim and finally goes off. Continue charging for about 2 hours, then the battery will have been fully charged.
Page 11
waveform. In this case, the output waveform is precise measurement can be measured. A more shifted with respect to the leading edge of the precise measurement can be obtained if the Tp por- reference frequency pulse by a time interval cor- tion of the waveform is expanded horizontally.
Page 12
Fig. 8 Typical digital circuit using several time-related waveforms...
Page 13
waveforms No. 4 through No. 8 and No. 10, would any of the waveforms may be displayed without be displayed on Channel 2 although other timing readjustment T R I G L E V E L control. comparisons may be desired. Waveforms No.
Page 14
Distortion M e a s u r e m e n t : 7. Adjust C H 1 and C H 2 vertical sensitivity con- trols ( V O L T S / D I V and V A R I A B L E ) so that the A n amplifier stage, or an entire amplifier unit, may waveforms are as large as possible without ex- be tested for distortion with this oscilloscope.
Page 15
can be examined more closely either by increasing Fig. 12 demonstrates the typical oscilloscope set- the sweep time or by pulling the POSITION tings a s well as the basic test circuit. Typical input control to obtain 5 times magnification. and output waveforms are shown on the os- cilloscope display.
Page 16
S t e r e o Amplifier S e r v i c i n g : Amplifier Phase Shift Measurements: Another convenient use for dual-trace oscilloscope Phase measurements can be made by several is in troubleshooting stereo amplifiers. If identical methods using oscilloscope.
Page 17
shown in Fig. 15B. The distance between cor- cilloscope control setting for viewing the alternate responding points on the horizontal axis for the two V I T S . w a v e f o r m s then represents the phase shift Most network television signals contain built-in...
Page 18
signal, followed by sine wave frequencies of 0.5 MHz, 1.0 MHz, 2.0 MHz, 3.0 MHz, 4 . 0 MHz and 3.58 MHz. This sequence of frequencies is called the "multi-burst". This multi-burst portion of the V I T S is the portion that can be most valuable to the technician.
Page 19
then our reasoning is right, and the IF amplifier the control in a counterclockwise direction requires realignment. If the poor response at 2 moving the traces to the left until the expand- ed information appears a s shown in Fig. 16. Because of the low repetition rate and the high the tuner input is misadjusted, causing a bite on only one channel.
Page 20
Composite Video Waveform Analysis normal signals. This method (signal-tracing) may be used to trace a signal by measuring several Probably the most important waveform in televi- points in the signal path. measurements sion and video servicing is the composite waveform proceed along the signal path, a point may be consisting of the video signal, the blanking pedestal found where the signal disappears.
Page 21
1. Tune the television set or video recorder blanking pulses are negative, use the " — " receiver to a local channel. position. A test tape or a signal generator also can be 13. Push in the T R I G L E V E L control and rotate to a used for service work.
Page 22
FM R E C E I V E R A D J U S T M E N T S 5. Set the marker generator precisely to 10.7 MHz. The marker "pip" should be in the Procedure: center of the bandpass. 1.
Page 23
A U D I O S I G N A L G E N E R A T O R V E R T I C A L A U D I O H O R I Z l O N T A L S I Z E S I Z E N E T W O R K B E I N G...
Page 24
Frequency M e a s u r e m e n t : tains a large number of odd harmonics. By injec- ting a 5 0 0 Hz sine w a v e into an amplifier, w e can 1. Connect the sine wave of known frequency to evaluate amplifier response at 5 0 0 Hz only, but by the C H 2 input jack of the oscilloscope and set injecting a square wave of the same frequency w e...
Page 25
venient viewing height. of A G - G N D - D C switch. However, the A C 5. Adjust the s w e e p time controls for one cycle of position may be used without affecting the square wave display on the screen. result except at very low frequency (below 5 6.
Page 26
In a typicaf wide band amplifier, a square wave check reveals many distortion characteristics of the circuit. The response of an amplifier is indicated in Fig. 20. revealing poor low-frequency response along with the overcompensated high-frequency boost. A 1 0 0 Hz square w a v e applied to the input of this amplifier will appear as in Fig.
Page 27
practically infinite number harmonic com- a leading direction. F i g . 3 5 indicates a 1 0 ° phase ponents. If an abnormal rise in amplifier response shift in a low frequency component in a lagging occurs at gigh frequencies, the high frequency direction.
Page 28
Fig. 37 indicates high frequency boost in an amplifier accompanied by a lightly damped " s h o c k " transient. The sinusoidal type of diminishing oscillation along the top of the square w a v e indicates a transient os- cillation in a relatively high "...
P R E C A U T I O N S 7. Installation of oscilloscope 1. Do not use the unit in any of the following The carrying handle of the unit turns 9 0 degree locations: angle in either direction (See Fig. 40), permit- * Places where the unit is exposed to direct ting the unit to be placed horizontally, vertical- sunlight.
Page 30
Caution: 3. Adjustment of Cut-off Voltage Connect an external DC power (12V) to the Before removing the case, be sure to turn off the battery connecting leads. Adjust V R 5 0 2 on power. Note that a high voltage ( 1 5 0 0 V max) is the power supply circuit board so that the present on the C R T socket and the printed circuit power lamp goes off when the voltage is...
Page 31
H O R I Z O N T A L C I R C U I T P O W E R S U P P L Y C I R C U I T B O A R D B O A R D ( T O P V I E W ) ( B O T T O M...